icon
An official government website affiliated with the Government of the Kingdom of Saudi Arabia.
icon Thu 25 Jun 2026
Beta version
icon
Official Saudi government website domains end with gov.sa

All official websites of government entities in the Kingdom of Saudi Arabia end with.gov.sa.

icon
Government websites use the protocol. HTTPS for encryption and security.

All official websites of government entities in the Kingdom of Saudi Arabia end with.gov.sa.

icon
Registered with the Digital Government Authority under number:
20250831619
Suggestions

The Saudi Food and Drug Authority would like to warn all healthcare practitioners within healthcare facilities from using Breathing Circuits used with Life Pulse High-Frequency Ventilators manufactured by Bunnell

Medical Devices

The Saudi Food and Drug Authority would like to warn all healthcare practitioners within healthcare facilities from using Breathing Circuits used with Life Pulse High-Frequency Ventilators manufactured by Bunnell

icon التاريخ: 2013-01-27
قسم: Medical Devices
نوع المواضيع: General Health

The Saudi Food and Drug Authority would like to warn all healthcare practitioners from using some affected lot ( attached ) of  Breathing Circuits used with the Life Pulse High-Frequency Ventilator manufactured by Bunnell. 

SFDA has become aware through the international regulatory resources, that the heater wire insulation of the breathing circuit - Disposable Patient Circuits-  manufactured by Bunnell- can melt, which  may cause a shortage that creates sparks and smoke in the circuit near the humidifier cartridge , leading to potential adverse events of  burns ,  injuries, lung infections or low oxygen levels in the patient’s blood – god forbid -

Saudi Food and Drug Authority encourages that all healthcare practitioners to adhere to the following :

- Identify any affected  product in your inventory and contact Saudi Food and Drug Authority .

- If you choose to use the affected product, Bunnell recommends that you perform the following actions to reduce the likelihood of this problem occurring :

1-  Visually inspect the circuit and the internal heating wire for thermal damage  every 30-60 min (e.g., sections of exposed wire without insulation, melting, or substantial discoloration). In particular, examine the red wires near the humidification chamber.

2-  Replace the circuit if a "Circuit Low Temp" alarm unexpectedly activates and the low temperature condition is not resolved within 5 min after following appropriate troubleshooting as described in the operator's manual.

3-  Replace the breathing circuit if displayed temperature is noted to be below the set temperature for >5 min.

 
 
Affected lot No’s
Catalog No. 937; Lot Nos.: 12C092, 12C102, 12C115, 12C125, 12C136, 12D159, 12D172, 12D189, 12E204, 12E211, 12E224, 12F241, 12F254, 12F271, 12G279, 12G290, 12G307, 12G321, 12H330, 12H349, 12I362, 12I371
 

 

 


 

 

 

 

Saudi Food and Drug Authority encourages all  to report any adverse event associated with medical devices through the National Center for Medical Devices Reporting :

for more information please visit the Saudi Food and Drug Authority website :

Media 2069